Carsan Engineering, Inc.
221 Corporate Circle
Suite H
Golden, CO 80401
(303) 237-9608
|
Carsan Engineering, Inc.
CEP300AHX, CEP300BHX,
CEP300CHX
300W Mercury-Xenon
Power Supply Specification
The CEP300 power supply is designed to run mercury-xenon
arc lamps in a constant-power mode. The output power is adjustable from
180 to 315 Watts with a built-in potentiometer. EMI line-filtering is
built-in to the unit. The supply includes an isolated +12V output for
powering external fans or electronics. The CEP300 power supply is designed
to switch to constant current mode by removing two easily accessible shorting
jumpers. Active Power Factor Correction to meet EMC limits for harmonic
current emissions, and limitations of voltage fluctuations and flicker.
Output power: 180 - 315 Watts, constant power
Output voltage compliance: 12.0 to 30.0 V operating, >110 VDC
during ignition.
Output regulation: output power held to within +/- 5% over all
input, output, and environmental conditions.
|



|
Output current: 8.0 to 18.0
Adc
Output ripple: <5%*
Efficiency: >70% at 300 W output, 120 VAC input
Thermal protection: ballast is disabled when temperature
exceeds 90 deg C.Unit will automatically restart after cooling down.
Isolated Auxiliary output (SELV Rated): +12.5VDC±2%,
2.0 A max.
Optically Isolated Status & Control Connector (SELV rated):
- Remote Enable
- Lamp Lit Status
- Linear-response remote intensity control
- Lamp over-voltage/under-voltage indicator
| Ground Leakage: |
AHX: <300 µA |
| BHX: <10 µA |
| CHX: <100 µA |
|
Regulatory Compliance: Approved to UL2601 with CB scheme
(E177225). Complies with EN55011 Class B Emissions. Meets EN 61000-3-2
and EN61000-3-3. CE-marked.
|
Line Input:
|
100 - 240 VAC, ± 10%, 50-60 Hz,
6.5 Arms max. |
|
Environmental:
|
0 deg C to 45 deg C operating. |
|
Altitude:
|
-1000 ft. to 12,000 ft. MSL. |
|
Weight:
|
2.7 lbs (1.2 Kg). |
|
Dimensions:
|
6.50" x 4.65",
2.60" tall (165mm x 114mm, 66mm tall). |
|
Ignitor:
|
25 KV ignition spike. Positive-side
ignition. Ignitor is removable.
Minimum repetition rate is 10 strikes/second at 90VAC.
Ignition pulses will continue for 6 ± 1 seconds. This feature
may be disabled via jumper, in which case the ignition pulses will
continue until the lamp ignites. |
|
Ignitor:
|
>100,000 strikes |
* ripple is measured in a DC to 20 MHz bandwidth.
Specifications Subject
to change without notice.
| Product # 91444-00002 |
Revision C
|
|