Carsan Engineering, Inc.
221 Corporate Circle
Suite H
Golden, CO 80401
(303) 237-9608
|
Carsan Engineering, Inc.
CEP100AHG, CEP100BHG,
CEP100CHG
Mercury Power Supply
Specification
The CEP100 power supply is designed to run mercury arc lamps in a constant-power mode. The output power is
adjustable from 90 to 135 Watts via a potentiometer on the circuit board. Output power can also be controlled from an
external voltage input. EMI line-filtering is built-in to the unit. The supply includes an isolated low voltage power supply with
optional multiple outputs.
Output power: 90 - 135 Watts, constant power
Output voltage compliance: 18.0 to 30.0 V operating.
Output regulation: output power held to within +/- 5% over all
input, output, and environmental conditions.
|



|
Output current: 2.0 to 6.5 Adc (8.0A during run-up)
Low-frequency output ripple: <2.0%„
Broadband output ripple: <0.2 Ap-p*
Efficiency: >70% at 110 W output, 120 VAC input
Thermal protection: ballast is disabled when heatsink temperature
exceeds 90 deg C. Unit will automatically restart after cooling
down.
Isolated Auxiliary output (SELV-rated):
|
Option
|
+12V,2.5A
|
+24V,1.0A
|
+5V,4.0A
|
+12V,0.3A
|
|
000
|
ü
|
×
|
×
|
×
|
|
001
|
ü
|
×
|
ü
|
×
|
|
002
|
ü
|
×
|
ü
|
ü
|
|
100
|
×
|
ü
|
×
|
×
|
|
101
|
×
|
ü
|
ü
|
×
|
|
|
Optically Isolated Status & Control
Connector (UL-rated SELV circuit)
- Remote Enable
- Lamp Lit Status
- Lamp Intensity Control Input
|
| Ground Leakage: |
AHG: <300 µA |
| BHG: <10 µA |
| CHG: <100 µA |
|
Regulatory Compliance: Medical Approval to EN60601-1 and TÜV Mark. Complies with EN55022. Meets EN
91000-3-2 and EN61000-3-3.
|
Line Input:
|
100 - 240 VAC, +/- 10%, 50/60 Hz, 3.2 Arms max. |
|
Environmental:
|
0 deg C to 50 deg C operating. |
|
Altitude:
|
-1000 ft. to 12,000 ft. MSL. |
|
Weight:
|
1.5 lbs (0.68 Kg). |
|
Dimensions:
|
6.50" x 4.50",
1.65" tall (165mm x 114mm, 42mm tall). |
|
Ignitor:
|
16 - 25 KV ignition spike, negative ignition.
Minimum repetition rate is 0.8 strikes/second at 90 VAC. Typical repetition
rate is 1.2 strikes/second (120 VAC).
Ignition pulses will continue until lamp ignites. |
„ Low frequency ripple is measured in a DC to 1 KHz bandwidth
and stated as percent of DC output curren
Specifications Subject
to change without notice.
| Product # 91456-00001 |
Revision D
|
|