Carsan Engineering, Inc.
CE300AHX, CE300BHX,
and CE300CHX
Mercury-Xenon Power
Supply Specification
The CE300 power supply is designed to run mercury-xenon arc lamps in a constant-power mode. The output power is
adjustable from 270-315 Watts with a built-in potentiometer. EMI line filtering is built-in to the unit. The CE300BHX
has smaller Y-capacitors to meet the 100 µA leakage requirement for many U.S. hospitals.
Output power: 270 - 315 Watts, constant power
Output voltage compliance: 20.0 to 25.0 V operating, >110 VDC.
during ignition.
Output regulation: output power held to within +/- 5% over all
input, output, and environmental conditions.
|



|
Output current: 13.0 to 18.0 Adc
Output ripple: <0.7 Ap-p *
Efficiency: >70% at 300 W output, 120 VAC input
Thermal protection: ballast is disabled when temperature
exceeds 90 deg C.Unit will automatically restart after cooling down.
Isolated Auxiliary output (UL-rated): +12V fan power, 800
mA max.
Optically Isolated Status & Control Connector (UL-rated
circuit)
- Remote Enable
- Lamp Lit Status
| Ground Leakage: |
AXE: <300 µA
BXE, <10 µA
CXE, <100 µA
|
|
Regulatory Compliance: Approved to UL2601 (E177225). Complies with EN55011 Class B Emissions. CE-marked.
|
Line Input:
|
100 - 240 VAC, +/- 10%, 47 - 63 Hz, 6.5
Arms max. |
|
Environmental:
|
0 deg C to 50 deg C operating. |
|
Altitude:
|
-1000 ft. to 12,000 ft. MSL. |
|
Weight:
|
2.5 lbs (1.14 Kg). |
|
Dimensions:
|
6.50" x 4.65",
2.80" tall (165mm x 118mm, 71mm tall). |
|
Ignitor:
|
25 KV ignition spike. Negative-side
ignition.
Minimum repetition rate is 10 strikes/second at all input voltages.
Ignition pulses will continue until lamp ignites.
Ignitor Life: >500,000 strikes.
|
* ripple is measured in a DC to 20 MHz bandwidth.
Specifications Subject
to change without notice.
| Product # 91434-00002 |
Revision C
|
|