Carsan Engineering, Inc.
221 Corporate Circle
Suite H
Golden, CO 80401
(303) 237-9608
|
Carsan Engineering, Inc.
CE105AHX, CE105BHX,
and CE105CHX
Mercury-Xenon Power
Supply Specification
The CE105 power supply is designed to run xenon arc lamps in a constant-current mode. The output current is
adjustable from 5.0 to 8.0 Amps via a potentiometer on the circuit board. EMI line-filtering is built-in to the unit. The
CE105BXE is the same as the A-model except it does not have Y-capacitors in the line filter section. The CE105CXE
has smaller Y-capacitors to meet the 100 ~uA leakage requirement of many U.S. hospitals.
Output power: 62 - 144 Watts, constant current
Output voltage compliance: 9.5 to 16.0 V operating, >110 VDC
during ignition.
Output regulation: output current held to within +/- 5% over all
input, output, and environmental conditions.
|



|
Output current: 5.0 to 8.0 Adc
Output ripple: <0.4 Ap-p *
Efficiency: >72% at 144 W output, 120 VAC input
Thermal protection: ballast is disabled when temperature
exceeds 90 deg C. Unit will automatically restart after cooling
down.
Auxiliary output (non-isolated): +12V fan power, 650 mA max.
12.OV to 12.8V at 250mA load.
Optically Isolated Status & Control Connector (UL-rated
circuit)
- Remote Enable
- Lamp Lit Status
- Remote Lamp Intensity Control
| Ground Leakage: |
AXE:
|
<300 uA at 136 VAC, <500
uA at 265 VAC |
| |
BXE:
|
<10 uA at all input voltages |
| |
CXE:
|
<100 uA at 136 VAC, <200
gA at 265 VAC |
|
Regulatory Compliance: Approved to UL2601 (E177225). Complies with EN55011 Class B Emissions. CE-marked.
|
Line Input:
|
100 - 240 VAC, +/- 10%, 47 - 63 Hz, 2.1
Arms max. |
|
Input Surge Current:
|
<30 Amps peak at turn-on, for all
input voltages. |
|
Environmental:
|
0 deg C to 45 deg C operating. |
|
Altitude:
|
-1000 ft. to 12,000 ft. MSL. |
|
Weight:
|
2.5 lbs (1.14 Kg). |
|
Dimensions:
|
6.00" x 4.15",
2.80" tall (152mm x 105mm, 71mm tall). |
|
Ignitor:
|
25 - 30 KV ignition spike.
Negative-side ignition.
Minimum repetition rate is 0.8 strikes/second at 90 VAC. Typical repetition
rate is 1.2 strikes/second (120 VAC).
Ignition pulses will continue until lamp ignites.
|
* ripple is measured in a DC to 20 MHz bandwidth.
Specifications Subject
to change without notice.
| Product # 91435-00003 |
Revision C
|
|